Wöhler A 550 NO Flue Gas Analyzer (Standard set)

Flue Gas Analyzer WÖHLER A 550 (Standard set) Measure O2+CO 4,000 ppm+NO 3,000 ppm, Oxygen range 0.0 – 21.0 %, 3.7 V 5800 mAh battery power

Description

Flue Gas Analyzer WÖHLER A 550 NO (Standard set)

As robust as cast iron, but as easy to operate as a smartphone – that’s what Wöhler’s new flue gas analyzer feels like in your hand. Appropriately, the advantages are equally substantial: A tip with your finger on the Wöhler A 550 is all that is required to perform analyses, inspections and settings on residential and commercial appliances. The flue gas analyzer is equipped with coded connections for the 4 Pa test and even »NEW AND UNIQUE« a wood moisture content measurement function to VDI Guideline 4206 (VDI = Association of German Engineers).

Wöhler, Germany

Feature

  • Measure gas, O2 + CO 4,000 ppm + NO 3,000 ppm
  • 7 meters long
  • USB charging head and Micro-USB charging cable
  • Have filters to filter water
  • Has a rough filter to filter large dust
  • There are 25 filter wadding filters.
  • save 1,000 data

STANDARD EQUIPMENT

  • – Wöhler A 550 Flue Gas Analyzer with Bluetooth, USB and IR-interface
    – O2 sensor
    – CO sensor 4,000 ppm
    – CO protection pump
    – NO sensor 3,000 ppm
    – cable hose 1.7 m
    – changeable gas probe 295 mm
    – combustion air temp. plug
    – USB-charger with Micro-USB cable
    – Lithium-ion rechargeable battery
    – 1 waterstop filter
    – 1 coarse filter
    – 25 wadding filters
    – plastic case

General Information
  • 3.7 V, 5800mAh battery power
  • Battery operating time : Approx. 12 h (depends on operating status and display illumination)
  • Storage temperature : -4 °F to +122 °F
  • Operating temperature : +41 – 104 °F to maintain stated accuracy
  • Weight : 2.75 lbs
  • Dimensions : 8.7" x 6.3" x 2.2" (without probe)
  • Length of cable-hose : 67"
Technical information
Oxygen concentration (O2) in flue gas
  • Display : Volume % referenced to dry flue gas
  • Measurement principle : Electrochemical sensor
  • Range 0.0 - 21.0 %
  • Accuracy ±0.3 %
Carbon monoxide (CO 4,000 ppm
  • Range 0 - 4,000 ppm
  • Resolution 1 ppm
  • Accuracy ±20 ppm (< 400 ppm), otherwise 5 % of reading
Nitric oxide concentration (NO)
  • Range 0 - 3,000 ppm (continuously up to 1,000)
  • Resolution 1 ppm
  • Accuracy ±5 ppm (< 100 ppm), otherwise 5 % of reading
Nitrogen dioxide concentration (NO2) in flue gas (optional)
  • Range 0 - 1000 ppm (continuously up to 200)
  • Resolution 1 ppm
  • Accuracy ±5 vol. ppm (< 100 ppm), otherwise 5 % of reading
Sulfur dioxide (SO2) (optional)
  • Range 0 - 5,000 ppm
  • Resolution 1 ppm
  • Accuracy ±10 vol. ppm (0 - 200 ppm), otherwise 5 % of reading
Chimney draft / differential pressure (PD)
  • Range 0.00 ± 44.16 inWC
  • Resolution 0.004 inWC (< 4.01 inWC), otherwise 0.004 inWC, with ventilation loss measurement 0.004 inWC
  • Accuracy 0.0012 inWC (< 0.04 inWC), otherwise 3 % of reading, Drift < 0.00081 inWC in 5 minutes
Flue gas temperature (Ts)
  • Range -4 °F to 1,472 °F
  • Resolution 32.18 °F
  • Accuracy 32 - 271.4 °F: ±35.6 °F, 271.4 - 1,472 °F: ±1.5 % of reading
Combustion air temperature (Ta)
  • Range -4 °F to 212 °F
  • Resolution 32.2 °F
  • Accuracy ±33.8 °F
Wood moisture
  • Range 10.0 – 40.0 %
  • Resolution 0.1 %
  • Accuracy ±40 % of reading

เอกสารข้อมูลผลิตภัณฑ์

Download File : woehler A550 technical data

เอกสารคู่มือ

Download File : Woehle A550 EN MANUAL