WÖHLER A 550 L NO Flue Gas Analyzer (Standard set)

Code: 8148

Flue Gas Analyzer Measure O2+CO 35,000 ppm+NO 3000 ppm , 3.7 V 5800mAh battery power , Have filters to filter water&dust

Description

Flue Gas Analyzer WÖHLER A 550 L NO (Standard set)

As robust as cast iron, but as easy to operate as a smartphone – that’s what Wöhler’s new flue gas analyzer feels like in your hand. Appropriately, the advantages are equally substantial: A tip with your finger on the Wöhler A 550 is all that is required to perform analyses, inspections and settings on residential and commercial appliances. The flue gas analyzer is equipped with coded connections for the 4 Pa test and even »NEW AND UNIQUE« a wood moisture content measurement function to VDI Guideline 4206 (VDI = Association of German Engineers).

Wöhler, Germany

Feature

  • Measure gas, O2 and CO 35,000 ppm, NO 3,000 ppm
  • 7 meters long
  • USB charging head and Micro-USB charging cable
  • Have filters to filter water
  • Has a rough filter to filter large dust
  • There are 25 filter wadding filters.
STANDARD EQUIPMENT
  • Wöhler A 550 L Flue Gas Analyzer with O2 + CO 35.000 ppm + NO 3.000 ppm
  • cable hose 1.7 m
  • gas probe 275 mm
  • USB-charger with Micro-USB cable
  • Lithium-ion rechargeable battery
  • 1 waterstop filter
  • 1 coarse filter
  • 25 wadding filters
  • plastic case

General Information
  • 3.7 V, 5800mAh battery power
  • Battery operating time : Approx. 12 h (depends on operating status and display illumination)
  • Storage temperature : -4 °F to +122 °F
  • Operating temperature : +41 – 104 °F to maintain stated accuracy
  • Weight : 2.75 lbs
  • Dimensions : 8.7" x 6.3" x 2.2" (without probe)
  • Length of cable-hose : 67"
Technical information
Oxygen concentration (O2) in flue gas
  • Display : Volume % referenced to dry flue gas
  • Measurement principle : Electrochemical sensor
  • Range : 0.0-21.0 vol. %
  • Accuracy : ± 0.3 Vol.-%
Nitrogen dioxide concentration (NO2) in flue gas (optional)
  • Display : Volume ppm referenced to dry flue gas
  • Measurement principle : Electrochemical sensor
  • Range : 0 – 1000 vol. ppm (continuously up to 200); resolution 1 vol. ppm
  • Accuracy : ±5 vol. ppm (< 100 ppm), otherwise 5 % of reading
Carbon monoxide (CO 35,000 ppm
  • Range 0 - 35,000 ppm
  • Resolution 1 ppm
  • Accuracy ±100 ppm (< 1,000 ppm),
    otherwise 10 % of reading (with H2 < 5% of reading)
Nitric oxide concentration (NO)
  • Range 0 - 3,000 ppm (continuously up to 1,000)
  • Resolution 1 ppm
  • Accuracy ±5 ppm (< 100 ppm), otherwise 5 % of reading
Sulfur dioxide (SO2) (optional)
  • Range 0 - 5,000 ppm
  • Resolution 1 ppmAccuracy ±10 vol. ppm (0 - 200 ppm), otherwise 5 % of reading
Chimney draft / differential pressure (PD)
  • Range 0.00 to ± 44.16 inWC
  • Resolution 0.004 inWC
  • Accuracy 0.008 inWC (< 0.16 inWC), otherwise 5 % of reading
Flue gas temperature (Ts)
  • Thermocouple measurement principle (NiCr-Ni)
  • Range measurements -20 to 100 °C
  • Resolution 0.1 °C
  • Accuracy ±1 °C
Combustion air temperature (Ta)
  • Thermocouple measurement principle (NiCr-Ni)
  • Range measurements -20 to 800 °C
  • Resolution 0.1 °C
  • Accuracy 0 to 133 °C (±2 °C), 133 to 800 °C (±1.5 %)
ความชื้นของเนื้อไม้
  • สามารถอ่านค่าออกมาเป็นมวลของน้ำ (mass)
  • หลักการวัดแบบ Electrical resistance measurement
  • ช่วงของการวัด 10.0 ถึง 40.0 %
  • ค่าความละเอียด 0.1 %
  • ค่าความถูกต้อง ±40 % (ทดสอบกับ VDI 4206 Part 4)

 

เอกสารข้อมูลผลิตภัณฑ์

Download File : Wöhler A 550 series data