WÖHLER A 450 H Flue Gas Analyzer 10,000 ppm (Standard set)

Code: 7165

Flue Gas Analyzer 10,000 PPM (H2 compensated), Measurement of O2, COV, NOx from chimneys in various industries, Measuring range COV 0 to 10,000 ppm

Description

WÖHLER A 450 H Flue Gas Analyzer (Basic Unit) (Standard set)

This is where Wöhler excels: A reasonably priced flue gas analyzer that offers all measurement and tuning functions. A small and lightweight flue gas analyzer enclosed in an eye-catching, sturdy housing. The Wöhler A 450 is a handheld device with user friendlytouchscreen control, displays all measurement data on screen or connect and view on your smart devices.

Wöhler, Germany

Feature

  • Open the meter immediately
  • Durable and compact measuring instrument for measurement and built-in adjustment function
  • 5-inch color touch screen, showing 14 values on one page
  • Wireless network connection
  • Wohler A 450 app for both Android and iOS

More information

  • Wöhler A 450 H with 10.000 ppm CO sensor (H2 compensated)
  • WIFI, USB and IR interface
  • probe 250 mm with 1.5 m hose
  • USB charger with micro USB cable
  • LI-ION battery 3,6 V / 6700 mAh
  • 25 cotton filters
  • Plastic case
  • cone Ø 8 mm

    5″ large color touchscreen
    All measured values available at a glance thanks to high-resolution display

    Interface to the printer
    Printout of the measured data on site

    Wöhler A 450 App
    Displaying the measured data on a mobile device

    Easy maintenance
    Condensate trap can easily be pulled out to change cotton wool and water stop fi lters.

    All values at a glance
    The tuning guide offers an advanced guidance to the technician to properly adjust the burner

    Perfect size
    Compact and light in a plastic case

General Information
  • Power supply :Lithium-Ion, rechargeable battery 3.7 V, 2,250 mAh, ชาร์จผ่าน USB
  • Operation time is about 17 hours (depending on usage status).
  • Charging time is about 7 hours.
  • Storage temperature -20 to 50 ° C
  • Operating Temperature 5 to 40 ° C
  • Weight 620 g (without sensor)
  • 160 x 110 x 45 mm (excluding probe)
  • 1.5 m long probe
Technical Information
Carbon monoxide (COV 10,000 ppm), H2 compensated

  • Range 0 - 5,000 ppm
  • Resolution 1 ppm
  • Accuracy ±20 ppm (< 400 ppm), rest 5 %
Oxygen concentration (O2)

  • Range 0.0 - 21.0 %
  • Accuracy ±0.3 %

 

Nitric oxide concentration (NOV)

  • Range 0 to 3,000 ppm; resolution 1 ppm
  • Accuracy ±5 ppm (< 100 ppm), otherwise 5 % of reading
Chimney draught / differential pressure (PD)

  • Range 0.00 ± 110.00 hPa
  • Resolution 1 Pa
  • Accuracy 2 Pa (< 40 Pa), otherwise 5 % of reading
Flue gas temperature (TS)

  • Range -20.0 to 800 °C
  • Resolution 0.1 °C
  • Accuracy 0 - 133 °C: ±2 °C, 133 - 800 °C: ±1.5 % of reading
Combustion air temperature (TA)

  • Range -20.0 to 100 °C
  • Resolution 0.1 °C
  • Accuracy ±1 °C
  • Battery operating time approx. 17 h