WTW – Thermoreactors CR 3200

Code: Thermoreactors CR 3200

Thermoreactors CR 3200 เตาให้ความร้อน สามารถทำอุณหภูมิได้ตั้งแต่ 25 ถึง 170 ◦C มีระบบทดสอบอุณหภูมิ ระหว่างอุณหภูมิเตาย่อย

Description

Thermoreactors CR 3200

Digestions of 24 round cells with pre-programmed and user-defined programs
In total 24 round cells (16 mm) in two seperate heating blocks using the same program
• 8 stored programs for easy and fast operation including COD fast digestion
• 8 user-defined programs from room temperature/25-170°C with high temperature digestion
• AQA/IQC with optional external temperature sensor for documentation

Programs for standard parameters COD including fast digestion (148°C, 20 min.), total nitrogen, total phosphorus, TOC and heavy metals of galvanizing with automatic heating and switch-off quick. User defined programs can be easily entered and stored from room temperature respectively 25-170°C to perform special digestions and thus allow high temperature digestion.
The very same program can be run for all vials at the same time; to perform two parallel digestion programs, the CR 4200 is the suitable choice. For the analytical quality assurance (AQA), the external temperature sensor TFK CR can be connected to the PC/printer instead of a round cell.

Programs for standard parameters
•1: 148 °C for 120 minutes
•2: 120 °C for 30 minutes
•3: 120 °C for 60 minutes
•4: 120 °C for 120 minutes
•5: 100 °C for 60 minutes
•6: 148 °C for 20 minutes
•7: 150 °C for 120 minutes
•8: 100 °C for 30 minutes

Key Features & Benefits:

  • สามารถทำอุณหภูมิได้ตั้งแต่ 25 ถึง 170 ◦C
  • ตั้งอุณหภูมิและเวลาการวัดได้ 8 โปรแกรม
  • มีระบบทดสอบอุณหภูมิ ระหว่างอุณหภูมิเตาย่อย และ Probe วัดอุณหภูมิ (Option)

SPECIFICATION
Reactor type Dry temperature control device with safety cover
Cell shafts 2 x 12 cell shafts for reaction cells, diameter: 16 ± 0.2 mm
Reaction time setting program settings or user-defined programs up to 180 min
Temperature setting stored program, user-defined settings from room temperature up to 170°C
Controlling accuracy ± 1 °C ± 1 digit
Temperature stability ± 0.5 K
Overtemperature protection 190 °C ± 5 °C
Dimensions D x W x H: 312 x 255 x 185 mm
Weight: 4 kg
Output unidirectional RS232 interface with 2 banana sockets for external temperature sensor (article-# 250 100), printer, PC
Power supply 115/230 VAC 50/60 Hz switchable via fuse setting
allowed voltage tolerance: ±15 % maximum power consumption: 560 W, Fuse 2x6.3 AT
Protection Class IP 20 according to EN 60529
Ambient temperature Operation +5 °C to +40 °C
Climatic class 2 according to VDI/VDE 3540
Relative humidity: Yearly mean: < 75 %
30 days /year: 95 % Other days: 85 % Light dew: yes
Site altitude: max. 5000 m above sea level
EMC EN 61326-1
FCC 47 CFR Part 15
Test certificates cETLus, CE
Safety standards EN 61010-1
EN 61010-2-010
UL 61010-1
CAN/CSA C22.2#61010-1
CAN/CSA C22.2#61010-2-010